product

Thermostatic cell for medical fluids

ENTH2000
Length mm
Height mm
Width mm
Generate PDF
 
The thermostatic cell is used for heating and maintaining body temperature of medical fluids for diagnostic and therapeutic use.
The thermostatic cell can be used for fluids contained in glass or plastic bottles, or in soft plastic bags.

- The thermostatic cell is heated using electrical resistance combined with forced air ventilation, operating at 220 VAC, 50Hz and 400 Watts.
- The device is made entirely of stainless steel both internally and externally, so it can be cleaned completely with normal cloths using ordinary detergents and disinfectants.
- Soaking or splashing with water or any fluid must be avoided to prevent damage to the device.
- The device is equipped with a temperature control system that maintains a temperature of 36°C/96,8°F with a tolerance of ± 1°C / 1,8°F.
- A constant temperature is assured, regardless of the ambient temperature and the temperature of incoming fluids, provided the ambient temperature is not above 35°C / 95°F or below 15°C / 59°F.
- This precise temperature is controlled and maintained thanks to a dual safety system and closedloop regulating system. A PID (proportional, integral, derived) corrective control system maximizes performance and sensitivity to disturbances such as ambient air temperature, container sizes and incoming container temperatures.

In order to prepare the fluid for ready use at the required time, the device is provided with a special user-programmable timer that automatically switches on the device at a pre-set time.

This programming must take place at the end of the day for the next day, or six hours and thirty minutes before the fluids are needed.

The device is also provided with a signal which can be viewed on the display and indicates when the temperature has been reached.

FRONT PANEL FUNCTIONS
- Ready
- Time
- Up time (programmed)
- Anomaly (error) signaling
The front panel switches allow the following functions to be programmed:
- Correct time-of-day
- Auto ON time to start heating

GENERAL SPECIFICATIONS
Temperature maintaining: 36°C/96,8°F
Power supply: 220 VAC 50 Hz.
Measurements: 540 x 380 x 515 mm (W,H,D)
Weight: 18 Kg
Capacity 27L
- 500 ml : 16 vials
- 250 ml : 20 vials
- 100 ml : 34 vials
Length mm
412,5
Width mm
390,5
Height mm
305
Additional Information
Dim int 412,5x390,5x305 mm -
Gross weight
20
Net weight
18
407950

We use cookies to ensure you get the best experience on our site. If you continue to use it, we will consider that you accept the use of cookies.